1.
Microsc Microanal ; 29(3): 994-1008, 2023 Jun 09.
Artigo em Inglês | MEDLINE | ID: mdl-37749665

RESUMO

We propose a modification of Wigner distribution deconvolution (WDD) to support live processing ptychography. Live processing allows to reconstruct and display the specimen transmission function gradually while diffraction patterns are acquired. For this purpose, we reformulate WDD and apply a dimensionality reduction technique that reduces memory consumption and increases processing speed. We show numerically that this approach maintains the reconstruction quality of specimen transfer functions as well as reduces computational complexity during acquisition processes. Although we only present the reconstruction for scanning transmission electron microscopy datasets, in general, the live processing algorithm we present in this paper can be applied to real-time ptychographic reconstruction for different fields of application.